In vitro comparison of nasal continuous positive airway pressure devices for neonates

86Citations
Citations of this article
33Readers
Mendeley users who have this article in their library.

Abstract

Objective: To compare the resistance in vitro of different devices used for the delivery of nasal continuous positive airway pressure (NCPAP) in neonates. Design: Flows of 4-8 litres/min were passed through a selection of neonatal NCPAP devices (single prong, Duotube, Argyle prong, Hudson prong, Infant Flow Driver), and the resultant fall in pressure measured using a calibrated pressure transducer. Results: The decrease in pressure (cm H2O) for each device (size in parentheses) at a constant flow of 6 litres/min was: Duotube: (2.5), 21; (3.0), 6.2; (3.5), 2.3; single prong: (2.5), 4.4; (3.0), 2.1; (3.5), 1.2; Argyle prong: (XS), 3.6; (S), 1.9; (L), 1.5; Hudson prong: (0), 3.1; (1), 1.8; (2), 0.6; (3), 0.4; (4), 0.3; Infant Flow Driver: (small), 0.3; (medium), -0.3; (large), -0.5. Conclusions: A large variation in the potential fall in pressure may occur in the clinical setting. Devices with short double prongs had the lowest resistance to flow. These results have implications in the selection of the optimal device/s for clinical application and for future comparisons in randomised trials of NCPAP in neonates.

Cite

CITATION STYLE

APA

De Paoli, A. G., Morley, C. J., Davis, P. G., Lau, R., & Hingeley, E. (2002). In vitro comparison of nasal continuous positive airway pressure devices for neonates. Archives of Disease in Childhood: Fetal and Neonatal Edition, 87(1). https://doi.org/10.1136/fn.87.1.f42

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free