Abstract
We present a method to calculate the electric (E)-fields within and surrounding a human body in a gradient coil, including E-fields induced by the changing magnetic fields and "conservative" E-fields originating with the scalar electrical potential in the coil windings. In agreement with previous numerical calculations, it is shown that magnetically-induced E-fields within the human body show no real concentration near the surface of the body, where nerve stimulation most often occurs. Both the magnetically-induced and conservative E-fields are shown to be considerably stronger just outside the human body than inside it, and under some circumstances the conservative E-fields just outside the body can be much larger than the magnetically-induced E-fields there. The order of gradient winding and the presence of conductive RF shield can greatly affect the conservative E-field distribution in these cases. Though the E-fields against the outer surface of the body are not commonly considered, understanding gradient E-fields may be important for reasons other than peripheral nerve stimulation (PNS), such as potential interaction with electrical equipment. © 2006 Wiley-Liss, Inc.
Author supplied keywords
Cite
CITATION STYLE
Mao, W., Chronik, B. A., Feldman, R. E., Smith, M. B., & Collins, C. M. (2006). Consideration of magnetically-induced and conservative electric fields within a loaded gradient coil. Magnetic Resonance in Medicine, 55(6), 1424–1432. https://doi.org/10.1002/mrm.20897