Finite-element study of the performance characteristics of an intradural spinal cord stimulator

4Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.
Get full text

Abstract

We have used finite-element (FE) modeling to investigate the mechanical compliance, positional stability and contact pressures associated with a novel type of spinal cord stimulator that is placed directly on the pial surface of the spinal cord in order to more selectively activate neural structures for relief of intractable pain. The properties used in the model are those of the actual prototype devices employed in recent in vitro and chronic in vivo tests. The agreement between predictions and experimental observations serves to validate our FE approach, which can now be used to further optimize the device's design and performance.

Cite

CITATION STYLE

APA

Grosland, N. M., Gillies, G. T., Shurig, R., Stoner, K., Viljoen, S., Dalm, B. D., … Howard, M. A. (2014). Finite-element study of the performance characteristics of an intradural spinal cord stimulator. Journal of Medical Devices, Transactions of the ASME, 8(4). https://doi.org/10.1115/1.4028421