Abstract
We describe the performance of a fine resolution neutron powder diffractometer, DN3, which has been installed at the neutron guide (NG2) in the Neutron Guide Hall about 71 m away from the Multipurpose G. A. Siwabessy reactor core. DN3 has a multi-detector system which consists of 32 3He neutron detectors and sollertype collimation. The resolution curve of the instrument was found to have little variation over a wide angular region. The best resolution is Δ2θ (FWHM) of 0.23 and Δd / d of 5×10-3, where 2q and d are scattering angle and dspacing, respectively. The instrument provides sample environment of low temperatures using a cryostat having a temperature range of 10-300K. Several experimental results, interlaboratory comparison study and he next activities to improve the performance of DN3 equipment have also been discussed.
Author supplied keywords
Cite
CITATION STYLE
Fajar, A., & Mugirahardjo, H. (2010). The performance of fine resolution neutron powder diffractometer at PTBIN-BATAN. Atom Indonesia, 36(1), 1–9. https://doi.org/10.17146/aij.2010.5