A 1-piece memory Nitinol intramedullary fixation device (Smart Toe; Stryker Corporation, Kalamazoo, MI) was used for proximal interphalangeal joint arthrodesis for correction of painful hammertoes in digits 2, 3, and 4. Sixty-five implants were placed in 35 patients. The mean age of our patients was 62.2 years (range = 27-82; standard deviation = 12.5). Mean follow-up time was 27 months (range = 12-40 months; standard deviation = 7). Overall, a 93.8% fusion rate was noted. Complications were noted in 4 patients (6.1%): 1 asymptomatic nonunion (1.5%), 2 hardware failures (3%), and 1 implant displacement (1.5%). To date, no patients required revisional surgery or hardware removal.Level of Evidence: V © 2013 The Author(s).

Cite

CITATION STYLE

APA

Sandhu, J. S., DeCarbo, W. T., & Hofbauer, M. H. (2013). Digital Arthrodesis With a One-Piece Memory Nitinol Intramedullary Fixation Device: A Retrospective Review. Foot and Ankle Specialist, 6(5), 364–366. https://doi.org/10.1177/1938640013496458