Abstract
Absolute calibration of a soft X-ray spectrograph has been performed using a white beam of synchrotron radiation. The calibrated spectrograph was a flat-field grazing-incidence spectrograph with an X-ray CCD detector for X-ray laser research. Absolute sensitivity of the spectrograph system can be obtained from transmitted spectra using filters made of several different materials, each providing an absorption-edge wavelength standard. The absolute sensitivity determined in this work shows nearly the same behaviour with wavelength as that in another calibration experiment using a laser-produced plasma as an X-ray source.
Author supplied keywords
Cite
CITATION STYLE
Fujikawa, C., Kawachi, T., Ando, K., Yamaguchi, N., & Hara, T. (1998). Absolute calibration of a soft X-ray spectrograph for X-ray laser research using white beam. Journal of Synchrotron Radiation, 5(3), 756–758. https://doi.org/10.1107/S0909049597016117