Abstract

Between December 2014 and March 2021, 144 patients with aortic (Ao) or mitral (Mi) paravalvular leaks (PVLs) were enrolled at 21 sites in 10 countries. Safety data were available for 137 patients, who were included in the safety analysis fraction (SAF), 93 patients with Mi PVLs and 44 patients with Ao PVLs. The full analysis set (FAS) comprised 112 patients with available stratum (aortic/mitral leak) as well as baseline (BL), 180-day or later assessments (2 years). Procedural success (implantation of the device with a proper closure of the PVL, defined as reduction in paravalvular regurgitation of ≥one grade as assessed by echocardiography post implantation) was achieved in 91.3% of FAS patients with Mi PVLs and in 90.0% of those with Ao PVLs. The proportion of patients suffering from significant or severe heart failure (HF), classified as New York Heart Association (NYHA) class III/IV, decreased from 80% at baseline to 14.1% at 2-year follow-up (FAS). The proportion of FAS patients needing hemolysis-related blood transfusion decreased from 35.5% to 3.8% and from 8.1% to 0% in Mi patients and Ao patients, respectively. In total, 35 serious adverse events (SAEs) were reported in 27 patients (19.7%) of the SAF population. The SAEs considered possibly or probably related to the device included device embolization (three patients), residual leak (two patients) and vascular complication (one patient). During follow-up, 12/137 (8.8%) patients died, but none of the deaths was considered to be device-related. Patients implanted with the Occlutech Paravalvular Leak Device (PLD) showed long-lasting improvements in clinical parameters, including NYHA class and a reduced dependency on hemolysis-related blood transfusions.

Cite

CITATION STYLE

APA

Onorato, E. M., Alamanni, F., Muratori, M., Smolka, G., Wojakowski, W., Pysz, P., … Bartorelli, A. L. (2022). Safety, Efficacy and Long-Term Outcomes of Patients Treated with the Occlutech Paravalvular Leak Device for Significant Paravalvular Regurgitation. Journal of Clinical Medicine, 11(7). https://doi.org/10.3390/jcm11071978

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free