Abstract
We report a case of 68-year-old female patient who suffered two Gamma 3 nail breakages following trochanteric fixation failure revision surgery. A patient was primarily treated with sliding hip screw. Revision surgeries of reverse trochanteric nonunion were performed with short Gamma 3 nails and bone grafting. The nail failures occurred 11 and 7 months after implantation.This undesired series of events were caused by inappropriate implant selection, thus replacing the nail to a modular total hip arthroplasty was more effective as salvage surgery.
Author supplied keywords
Cite
CITATION STYLE
Stepanovic, Z. L., & Ristic, B. M. (2015). Salvage total hip arthroplasty after two gamma 3 nail failures: A case report. Acta Chirurgiae Orthopaedicae et Traumatologiae Cechoslovaca, 82(3), 239–242. https://doi.org/10.55095/achot2015/036