Wirelessly Controlled Implantable System for On-demand and Pulsatile Insulin Administration

8Citations
Citations of this article
34Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We propose a wirelessly controlled implantable system for on-demand and pulsatile insulin delivery with a more convenient and safer strategy than currently available strategies. The system is a combined entity of a magnetically driven pump (i.e., an MDP), external control device (i.e., an ECD) and mobile app. The MDP for implantation consists of a plunger, barrel and drug reservoir, where an accurate amount of insulin can be infused in a pulsatile manner only at the time when a magnetic force is applied to actuate the plunger in the barrel. The ECD at the outside body can modulate the MDP actuation with an electromagnet and its control circuit, and this modulation can be wirelessly controlled by the mobile app. As a safety feature, the mobile app is programmed to pre-set the restrictions for the insulin dose and administration schedule to avoid overdose. The system is shown to infuse insulin in a highly reproducible manner, but it does not allow for insulin infusion when the pre-set restrictions are violated. When tested with diabetic rats, the profiles of insulin plasma concentration and blood glucose level are similar to those of animals treated with a subcutaneous injection of the same dose of insulin.

Cite

CITATION STYLE

APA

Lee, S. H., Ahn, J. W., Cho, Y. C., Kim, S. N., Lee, C., Ku, G. W., … Kim, H. C. (2019). Wirelessly Controlled Implantable System for On-demand and Pulsatile Insulin Administration. Scientific Reports, 9(1). https://doi.org/10.1038/s41598-019-41430-8

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free