A study of inflatable cuffs on endotracheal tubes: Pressures exerted on the trachea

21Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A simple method of measuring the pressure exerted by the cuff of an endotracheal tube on the trachea is described and has been used to measure the pressures exerted by 16 commercially available cuffs on the wall of a model trachea. The Shiley, Portex soft-seal, Kamen-Wilkinson (Bivona Fome) tubes had the lowest tracheal wall pressures. Using this method in vivo the changes in tracheal wall pressure exerted by a low-pressure cuff during percussion and vibration physiotherapy, and when the patient "fights the ventilator", were recorded. © 1976 Macmillan Journals Ltd.

Cite

CITATION STYLE

APA

Mackenzie, C. F., Klose, S., & Browne, D. R. G. (1976). A study of inflatable cuffs on endotracheal tubes: Pressures exerted on the trachea. British Journal of Anaesthesia, 48(2), 105–110. https://doi.org/10.1093/bja/48.2.105