Abstract
A mechanical heat switch used in calorimetry experiments down to 0.3 K is described and its performance analyzed in sufficient detail to be useful for design purposes. The thermal conductance of the indium-copper switch contacts was determined as a function of temperature and applied force over the range 0.3-2 K. These results are compared with other conductance measurements to evaluate choices of metals for heat switch contacts. © 1969 The American Institute of Physics.
Cite
CITATION STYLE
Colwell, J. H. (1969). The performance of a mechanical heat switch at low temperatures. Review of Scientific Instruments, 40(9), 1182–1186. https://doi.org/10.1063/1.1684194