Measurements outside resonance with piezoresponse force microscopy

3Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

In this work, we report the advantages of following a protocol to perform measurements of piezoresponse force microscopy (PFM) away from the tip-sample contact resonance. The protocol established in this work is based on Soergel's work [J. Phys. D: Appl. Phys. 44 (2011) 464003]. PFM measurements out-of-plane were carried out in the range of 10-100kHz. The experimental background, that affects the domain contrast during PFM measurements, was reduced by analyzing the X-signal from the lock-in amplifier (LIA) during a frequency sweep measurement on a glass slide (non-piezoelectric). Optimal contrast in X-signal was attained through set the phase of the LIA to reach a null contrast in the Y-signal. Then, the X-Y and R-Θ signals are stored using the optimal conditions established for LIA. The protocol described was applied to a sol-gel BiFeO3 film and a reference sample of periodically poled lithium niobate.

Cite

CITATION STYLE

APA

Garduño-Medina, A., Vázquez-Delgado, M. A., Diliegros-Godines, C. J., García-Vázquez, V., & Flores-Ruiz, F. J. (2021). Measurements outside resonance with piezoresponse force microscopy. In AIP Conference Proceedings (Vol. 2416). American Institute of Physics Inc. https://doi.org/10.1063/5.0068705