A Convenient RP-HPLC Method for Assay Bioactivities of Angiotensin I-Converting Enzyme Inhibitory Peptides

  • Wang W
  • Wang N
  • Zhang Y
  • et al.
N/ACitations
Citations of this article
44Readers
Mendeley users who have this article in their library.

Abstract

A convenient and accurate reversed-phase high-performance liquid chromatography (RP-HPLC) method for angiotensin I-converting enzyme inhibitory peptides assay was described in this paper. The mobile phase consisted of 70% A (0.05% TFA and 0.05% triethylamine in water, pH = 2.9–3.3) and 30% B (100% acetonitrile) using an Isogradient program. The flow rate was 0.5 mL/min. The absorb wavelength was 226.5 nm; the column temperature was controlled at 25°C. This method for angiotensin I-converting enzyme inhibitory peptides assay was convenient for the Iso-gradient program. The accuracy of the RT-HPLC method was verified by analyzing ACE inhibitory activity of the hydrolysate peptides of silkworm pupae protein, and the results showed that the RT-HPLC method was available for exploring new source of angiotensin I-converting enzyme inhibitory peptides rapidly and veraciously.

Cite

CITATION STYLE

APA

Wang, W., Wang, N., Zhang, Y., Cai, Z., Chen, Q., & He, G. (2013). A Convenient RP-HPLC Method for Assay Bioactivities of Angiotensin I-Converting Enzyme Inhibitory Peptides. ISRN Biotechnology, 2013, 1–5. https://doi.org/10.5402/2013/453910