Abstract

The ParaGraph is a new device for monitoring neuromuscular function using a piezoelectric motion sensor. In 20 patients, monitoring of neuromuscular block produced by cisatracurium 0.1 mg kg-1 was compared using the ParaGraph and a Relaxometer 2 mechanomyograph. The ParaGraph was quick to set up, and easy to operate and interpret. There were no significant differences in the time to 100% depression of T1/T0, time to 25% recovery of T1/T0 or time to recovery of T1/T0 from 25% to 75%, measured by the two monitors. When the difference between the two monitors was plotted against the average of the two measurements, the limits of agreement for T1/T0 (-42.95, +53.98%) and the train-of-four ratio, T4/T1 (-0.28, +0.21) were too wide to allow the values given by the two monitors for individual patients to be used interchangeably.

Cite

CITATION STYLE

APA

Dahaba, A. A., Klobucar, F. V., Rehak, P. H., & List, W. F. (1999). Comparison of a new piezoelectric train-of-four neuromuscular monitor, the ParaGraph, and the Relaxometer mechanomyograph. British Journal of Anaesthesia, 82(5), 780–782. https://doi.org/10.1093/bja/82.5.780