An SPR-Based In Situ Methane Sensor for the Aqueous and Gas Phase

5Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

This paper presents a fast, low-cost, precise methane sensor based on refractive index changes in a cryptophane A (CryptA)-doped polystyrene membrane. For the realization of this sensor, we built a surface plasmon resonance sensor with a refractive index resolution of 4.31 × 10-6 and investigated the optimal membrane thickness, i.e., a polymer layer of sufficient sensitivity with the lowest response time. For a membrane thickness of 760 nm, a limit of detection of 135 ppm and a response time constant of 45 s were found. Despite a comparable refractive index resolution and a higher CryptA content, the limit of detection is 3 orders of magnitude larger than that of a reported prototype. The sensor is capable of quantifying methane in the gas or aqueous phase. However, temperature, humidity, and ethanol vapor highly influence the signal. Although the principle of CryptA-based methane measurement is a promising, fast, and low-cost method, it will not be able to compete with state-of-the-art sensing in its current state.

Cite

CITATION STYLE

APA

Estelmann, A., Prien, R., Marz, W., Elbing, M., Harz, P., & Rehder, G. (2024). An SPR-Based In Situ Methane Sensor for the Aqueous and Gas Phase. Analytical Chemistry, 96(41), 16203–16214. https://doi.org/10.1021/acs.analchem.4c02875