Break of a stent delivery catheter and subsequent stent loss (SL) has been a rare event in the drug-eluting stent (DES) era. We here report a case of successful retrieval of a stent after a break if the delivery catheter and SL from a balloon catheter at a culprit lesion. We finally resolved this situation using a simple balloon technique for both the broken stent catheter inside of the guide catheter and the unexpanded stent in the culprit lesion. Thus balloons are an important weapon in our armamentarium in the cardiac catheterization laboratory for urgent retrieval of a lost stent. Their apt use definitely allowed our patient to avoid undergoing emergency cardiovascular thoracic surgery. Copyright © 2010 The Korean Society of Cardiology.