False Positive Transit Time Flowmetry Graft Failure in Multivessel Coronary Spasm following Off-Pump Coronary Artery Bypass Grafting

  • Kassimis G
  • Krasopoulos G
N/ACitations
Citations of this article
7Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Intraoperative Transit Time Flowmetry is currently recommended to assess graft patency during coronary artery bypass grafting (CABG), especially in presence of haemodynamic instability or inability to wean the patient from cardiopulmonary bypass, new regional wall motion abnormalities, or significant ventricular arrhythmias. The VeriQ™ system is one of the currently available systems, which detects imperfections that may be corrected by graft revision. In this case report, multivessel coronary spasm (CS) post-CABG interferes with these intraoperative parameters misleading initially into false results. Cardiac surgeons should bear in mind the limit of VeriQ in distinguishing between graft failure and CS. Angiography may be considered in patients with decreased graft flow despite revision of anastomosis and vasodilatory treatment for the definitive diagnosis.

Cite

CITATION STYLE

APA

Kassimis, G., & Krasopoulos, G. (2017). False Positive Transit Time Flowmetry Graft Failure in Multivessel Coronary Spasm following Off-Pump Coronary Artery Bypass Grafting. Case Reports in Cardiology, 2017, 1–3. https://doi.org/10.1155/2017/3186047