Abstract
Pulmonary vein isolation is the electrophysiological endpoint of complete conduction block at the level of the veno-atrial junction and must be explicitly distinguished from encircling PV ablation which frequently does not result in isolation. The prerequisites for successful PV isolation include a knowledge of the individual anatomy of the PVs and the left atrium, appropriate positioning of circular mapping catheters, and a knowledge of the electrophysiology of PV activation, in addition to effective ablation tools. Excessive ablation, and possibly complications, can be avoided by the recognition of non-PV myocardial contributions to electrograms recorded from within the PVs. The posterior wall of the left atrial appendage contributes far-field electrograms to recordings from all or nearly all left superior PVs, the low anterior left atrium to 80% of left inferior PV recordings and the superior vena cava to 23% of right superior PV recordings. Recognition of these far-field components is feasible and accurate in sinus rhythm as well as during ongoing atrial fibrillation. Finally, the creation of temporally stable PV isolation remains a currently unsolved problem although prolonged post isolation surveillance, may be helpful. © 2008, Japanese Heart Rhythm Society. All rights reserved.
Author supplied keywords
Cite
CITATION STYLE
Shah, D. (2008). Evaluating Pulmonary Vein Isolation. Journal of Arrhythmia, 24(4), 180–194. https://doi.org/10.4020/jhrs.24.180