Effect of bipolar electrode spacing on phrenic nerve stimulation and left ventricular pacing thresholds an acute canine study

22Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.

Abstract

Background-Phrenic nerve stimulation (PNS) is a common complication of cardiac resynchronization therapy when left ventricular (LV) pacing occurs via a coronary vein. The purpose of this study was to evaluate the effects of bipolar electrode spacing on PNS and LV pacing thresholds. Methods and Results-Electrophysiology catheters with standard (2 mm-5 mm-2 mm) or modified (1 mm-5 mm-1 mm) interelectrode spacing was, respectively, inserted in a posterior/lateral cardiac vein in a randomized order in 6 anesthetized dogs via jugular access. The phrenic nerve was dissected via a left minithoracotomy and repositioned over the vein as close as possible to one of the electrodes. The presence of PNS was verified (ie, PNS threshold <2 V at 0.5 ms in unipolar configuration). Bipolar pacing was delivered using the electrode closest to the phrenic nerve as the cathode, and multiple bipolar electrode spacing configurations were tested. During bipolar pacing, PNS threshold increased as bipolar electrode spacing was reduced (P<0.05), whereas LV pacing thresholds did not change significantly (P>0.05). Compared with a standard bipolar electrode spacing of 20 mm for LV leads, 1 and 2 mm bipolar electrode spacing resulted in a PNS threshold increase of 5.5±2.2 V (P=0.003) and 2.8±1.7 V (P<0.001), respectively. Similarly, PNS threshold increased by 6.5±3.7 V with 1 mm and by 3.8±1.9 V with 2 mm bipolar pacing (both P<0.001), compared with unipolar pacing. Conclusions-This study suggests that reducing LV bipolar electrode spacing from the standard 20 mm to 1 or 2 mm may significantly increase the PNS threshold without compromising LV pacing thresholds. © 2012 American Heart Association, Inc.

Cite

CITATION STYLE

APA

Biffi, M., Foerster, L., Eastman, W., Eggen, M., Grenz, N. A., Sommer, J., … Yang, Z. (2012). Effect of bipolar electrode spacing on phrenic nerve stimulation and left ventricular pacing thresholds an acute canine study. Circulation: Arrhythmia and Electrophysiology, 5(4), 815–820. https://doi.org/10.1161/CIRCEP.112.971317

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free