Abstract
We report the development and validation of a 12 parameter immunofluorescence flow cytometry method for the sensitive determination of cell concentrations, their expression of PD-1, and PD-1 receptor occupancy. Cell subsets include CD4+ and CD8+-T-cells, B-cells, natural killer cells, classical-, intermediate- and non-classical monocytes, and myeloid- and plasmacytoid dendritic cells. Cells were isolated from peripheral blood by density gradient centrifugation. The validation parameters included specificity, linearity, limit of quantification, precision, biological within- and between subject variations. The lower limit of quantification was 5.0% of PD-1+ cells. Samples were stable for at least 153 days of storage at −80°C. The clinical applicability of the method was demonstrated in 11 advanced cancer patients by the successful determination of immune cell concentrations, relative number of PD-1+ immune cells, and number of PD-1 molecules per immune cell. Shortly after infusion of nivolumab, receptor occupancy on CD8+-T-cells was 98%. Similar values were found predose cycle 2, suggesting receptor occupancy remained high throughout the entire cycle. © 2019 International Society for Advancement of Cytometry.
Author supplied keywords
Cite
CITATION STYLE
Pluim, D., Ros, W., Miedema, I. H. C., Beijnen, J. H., & Schellens, J. H. M. (2019). Multiparameter Flow Cytometry Assay for Quantification of Immune Cell Subsets, PD-1 Expression Levels and PD-1 Receptor Occupancy by Nivolumab and Pembrolizumab. Cytometry Part A, 95(10), 1053–1065. https://doi.org/10.1002/cyto.a.23873