Abstract
The recent increase in severe fever with thrombocytopenia syndrome (SFTS) cases has led to the development of the SFTS-QS kit (MiCoBioMed, Seongnam, Korea) for detecting the SFTS virus (SFTSV, now renamed Huaiyangshan banyangvirus). SFTS-QS is a qualitative real-time reverse transcription PCR assay based on lab-on-a- chip technology. We evaluated the performance of the SFTS-QS kit and compared it with that of the PowerChek SFTSV Real-time PCR kit (PowerChek; Kogene Biotech, Seoul, Korea). A total of 117 serum samples were simultaneously assayed using the SFTS-QS and PowerChek kits. Sanger sequencing targeting the S and M segments of SFTSV was performed as the reference method. The total turnaround time of the two kits was compared. The SFTS-QS results agreed with those of PowerChek with a kappa value of 0.92. The diagnostic sensitivity and specificity of the SFTS-QS kit were both 100% (14/14 and 103/103, respectively), whereas those of the PowerChek kit were 100% (14/14) and 98.1% (101/103), respectively. The results of SFTS-QS and PowerChek were comparable; however, the SFTS-QS kit required a shorter total turnaround time. The SFTS-QS kit produced accurate and fast results and thus could serve as a useful tool for detecting SFTSV.
Author supplied keywords
Cite
CITATION STYLE
Yoo, I. Y., Kim, J. Y., Yoon, Y. K., Huh, H. J., & Lee, N. Y. (2020). Comparison between the SFTS-QS Kit and the PowerChek SFTSV real-time PCR kit for the detection of severe fever with thrombocytopenia syndrome virus. Annals of Laboratory Medicine, 40(4), 317–320. https://doi.org/10.3343/alm.2020.40.4.317