In a head-to-head comparison of the MTBDRplus version 2.0 (Hain Lifescience), the Xpert MTB/RIF (Cepheid), and the Anyplex MTB/NTM (Seegene) assays, we demonstrated equal sensitivity (59/61; 96.7%) and specificity (53/54; 98.1%) for detecting rifampin resistance with further analysis of discordances. The Xpert assay does not detect isoniazid resistance while the Anyplex assay showed high false positivity.
CITATION STYLE
Strydom, K., Ismail, F., Matabane, M. M. Z., Onwuegbuna, O., Omar, S. V., & Ismail, N. (2015). Comparison of three commercial molecular assays for detection of rifampin and isoniazid resistance among Mycobacterium tuberculosis isolates in a high-HIV-prevalence setting. Journal of Clinical Microbiology, 53(9), 3032–3034. https://doi.org/10.1128/JCM.01691-15