Design, performance, and calibration of CMS hadron-barrel calorimeter wedges

32Citations
Citations of this article
42Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Extensive measurements have been made with pions, electrons and muons on four production wedges of the compact muon solenoid (CMS) hadron barrel (HB) calorimeter in the H2 beam line at CERN with particle momenta varying from 20 to 300 GeV/c. The time structure of the events was measured with the full chain of preproduction front-end electronics running at 34 MHz. Moving-wire radioactive source data were also collected for all scintillator layers in the HB. The energy dependent time slewing effect was measured and tuned for optimal performance. © 2008 Springer-Verlag / Società Italiana di Fisica.

Cite

CITATION STYLE

APA

Abdullin, S., Abramov, V., Acharya, B., Adams, M., Akchurin, N., Akgun, U., … Zeyrek, M. (2008). Design, performance, and calibration of CMS hadron-barrel calorimeter wedges. European Physical Journal C, 55(1), 159–171. https://doi.org/10.1140/epjc/s10052-008-0573-y