Rapid portal imaging with a high-efficiency, large field-of-view detector

50Citations
Citations of this article
25Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The design, construction, and performance evaluation of an electronic portal imaging device (EPID) are described. The EPID has the same imaging geometry as the current mirror-based systems except for the x-ray detection stage, where a two-dimensional (2D) array of 1 cm thick CsI(T1) detector elements are utilized. The ~18% x-ray quantum efficiency of the scintillation detector and its 30 X 40 cm2 field-of-view at the isocenter are greater than other area-imaging EPIDs. The imaging issues addressed are theoretical and measured signal-to-noise ratio, linearity of the imaging chain, influence of frame-summing on image quality and image calibration. Portal images of test objects and a humanoid phantom are used to measure the performance of the system. An image quality similar to the current devices is achieved but with a lower dose. With ~1 cGy dose delivered by a 6 MV beam, a 2 mm diam structure of 1.3% contrast and an 18 mm diam object of 0.125% contrast can be resolved without using image-enhancement methods. A spatial resolution of about 2 mm at the isocenter is demonstrated. The capability of the system to perform fast sequential imaging, synchronized with the radiation pulses, makes it suitable for patient motion studies and verification of intensity-modulated beams as well as its application in cone- beam megavoltage computed tomography.

Cite

CITATION STYLE

APA

Mosleh-Shirazi, M. A., Evans, P. M., Swindell, W., Symonds-Tayler, J. R. N., Webb, S., & Partridge, M. (1998). Rapid portal imaging with a high-efficiency, large field-of-view detector. Medical Physics, 25(12), 2333–2346. https://doi.org/10.1118/1.598443

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free