T-wave oversensing and inappropriate shocks: A case report

24Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A 27-year-old male with congenital long QT syndrome, SCN5A mutation experienced recurrent inappropriate exercise-related implantable cardioverter defibrillator (ICD) shocks. This device showed T-wave oversensing with double, which lead to these device discharges. Dynamic T-wave oversensing was reproducibly provoked at exercise treadmill testing and was confirmed as the mechanism leading to double counting. The insertion of a new pacing and sensing lead with increased R-wave amplitude did not solve the problem. Exchanging the existing ICD generator with one capable of automatic sensitivity control (Biotronik, Lexos DR, Biotronik, Berlin, Germany) completely eliminated T-wave oversensing and inappropriate shocks. © The Author 2008.

Cite

CITATION STYLE

APA

Srivathsan, K., Scott, L. R., & Altemose, G. T. (2008). T-wave oversensing and inappropriate shocks: A case report. Europace, 10(5), 552–555. https://doi.org/10.1093/europace/eun083