Abstract
An 84-year-old woman implanted with cardiac resynchronization therapy defibrillator underwent transvenous lead extraction 4 mo after the implant due to pocket infection. Atrial and right ventricular leads were easily extracted, while the attempt to remove the coronary sinus (CS) lead was unsuccessful. A few weeks later a new extraction procedure was performed in our center. A stepwise approach was used. Firstly, manual traction was unsuccessfully attempted, even with proper-sized locking stylet. Secondly, mechanical dilatation was used with a single inner sheath placed close to the CS ostium. Finally, a modified sub-selector sheath was successfully advanced over the electrode until it was free of the binding tissue. The post-extraction lead examination showed an unexpected fibrosis around the tip. No complications occurred during the postoperative course. Fibrous adhesions could be found in CS leads recently implanted requiring non-standard techniques for its transvenous extraction.
Cite
CITATION STYLE
Bontempi, L., Tempio, D., De Vito, R., Cerini, M., Salghetti, F., Dasseni, N., … Curnis, A. (2017). Unexpected challenging case of coronary sinus lead extraction. World Journal of Clinical Cases, 5(2), 46. https://doi.org/10.12998/wjcc.v5.i2.46