Protrusion of a loop-shaped 0.035-inch wire without the fracture of the self-expanding nitinol stent: A case report and experimental study

1Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We report a case of a loop-shaped 0.035-inch wire protruding through self-expanding nitinol stent struts. Our in vitro experiment suggests that, even if there are no stent fractures, the loop-shaped 0.035-inch wire has a potential to protrude through the struts of the self-expanding nitinol stents.

Cite

CITATION STYLE

APA

Karashima, E., Kawasaki, D., Yoda, S., Yasuda, S., & Kaneko, T. (2019). Protrusion of a loop-shaped 0.035-inch wire without the fracture of the self-expanding nitinol stent: A case report and experimental study. Clinical Case Reports, 7(2), 357–360. https://doi.org/10.1002/ccr3.1988