Prostate-specific antigen (PSA) is the established routine screening tool for the detection of early-stage prostate cancer. Given the laboratory-centric nature of the process, the development of a portable, ultra rapid high-throughput system for PSA screening is highly desirable. In this study, an advancedpoint-of-care system for PSA detection in human serum was developed based on a cellular biosensor where the cell membrane was modified by electroinserting a specific antibody against PSA. Thirty nine human serum samples were used for validation of this biosensory system for PSA detection. Samples were analyzed in parallel with a standard immunoradiometric assay (IRMA) and an established electrochemical immunoassay was used for comparison purposes. They were classified in three different PSA concentration ranges (0, <4 and ≥4 ng/mL). Cells membrane-engineered with 0.25 µg/mL anti-PSA antibody demonstrated a statistically lower response against the upper (≥4 ng/mL) PSA concentration range. In addition, the cell-based biosensor performed better than the immunosensor in terms of sensitivity and resolution against positive samples containing <4 ng/mL PSA. In spite of its preliminary, proof-of-concept stage of development, the cell-based biosensor could be used as aninitiative for the development of a fast, low-cost, and high-throughput POC screening system for PSA.
CITATION STYLE
Mavrikou, S., Moschopoulou, G., Zafeirakis, A., Kalogeropoulou, K., Giannakos, G., Skevis, A., & Kintzios, S. (2018). An ultra-rapid biosensory point-of-care (POC) assay for prostate-specific antigen (PSA) detection in human serum. Sensors (Switzerland), 18(11). https://doi.org/10.3390/s18113834