The aim of this study was to verify the drying effect on the reproducibility of DIAGNOdent (Dd) devices to detect caries-like lesions. Three areas were created in each of the 34 bovine incisors: sound (S), demineralized (DE) and remineralized (RE). One examiner measured each area with two Dd devices (denominated X and Y), twice under humid, and twice under dry condition. Intra-rater agreement according each device and inter-device agreement were estimated by kappa statistics (κ). Intra-rater agreement for device Y was substantial under humid (κ DE=0.68 and κ RE+S=0.68) and dry condition (κ DE=0.64 and κ RE+S=0.67). For device X, it was substantial under humid condition (κ DE=0.57 and κ RE+S=0.49), and it was almost perfect after air drying (κ DE=1.0 and κRE+S=1.0). Inter-device agreement was slight (κ =0.17) under humid condition, and it was substantial under dry condition (κ =0.62). As reproducibility increased under dry condition, drying is advised to detect caries-like lesions on free smooth surfaces when different devices are used.
CITATION STYLE
Pinelli, C., Loffredo, L. de C. M., & Serra, M. C. (2010). Effect of drying on the reproducibility of DIAGNOdent to detect caries-like lesions. Brazilian Dental Journal, 21(5), 405–410. https://doi.org/10.1590/S0103-64402010000500005