Abstract
Objectives: To estimate the minimum detectable iodine concentration on multiple dual-energy CT (DECT) platforms. Methods and materials: A phantom containing iodine concentrations ranging from 0 to 10 mg ml-1 was scanned with five dual-energy platforms (two rapid kilo volt switching (r-kVs), one dual source (DS), one sequential acquisition and one split-filter). Serial dilutions of 300 mg ml-1 iodinated contrast material were used to generate concentrations below 2 mg ml-1. Iodine density and virtual monoenergetic images were reviewed by three radiologists to determine the minimum visually detectable iodine concentration. Contrast-to-noise ratios (CNRs) were calculated. Results: 1 mg mL-1 (0.8 mg mL-1 corrected) was the minimum visually detectable concentration among the platforms and could be seen by all readers on the third-generation r-kVs and DS platforms. Conclusions: At low concentrations, CNR for monoenergetic images was highest on the DS platform and lowest in the sequential acquisition and split-filter platforms. Advances in knowledge: The results of this study corroborate previous in vivo estimates of iodine detection limits at DECT and provide a comparison for the performance of different DECT platforms at low iodine concentrations in vitro.
Cite
CITATION STYLE
Taylor, R. E., Mager, P., Yu, N. C., Katz, D. P., Brady, J. R., & Gupta, N. (2019). Iodine quantification and detectability thresholds among major dual-energy CT platforms. British Journal of Radiology, 92(1104). https://doi.org/10.1259/bjr.20190530