In-vitro myoglobin clearance by a novel sorbent system

15Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Rhabdomyolysis may lead to acute kidney injury following deposition of myoglobin in renal tubules. Although high-flux dialysis membranes may remove a substantial amount of myoglobin from plasma, this may still not be sufficient to prevent renal damage. We tested a new polymer sorbent, X-Sorb®, in vitro to determine its potential to clear myoglobin from solutions. Normal saline or human serum in which myoglobin was dissolved was perfused by a peristaltic pump through a column packed with the sorbent. After a 4-hour perfusion, the myoglobin level in normal saline fell from 200,000 ng/ml to virtually undetectable (<780 ng/ml). Perfusion through the sorbent was then found to lower concentrations of dissolved myoglobin in 3 different 110-ml samples of human serum consistently by > 90% over 4 hours. X-Sorb appears to be an effective sorbent for myoglobin and warrants a trial in vivo to determine whether it is equally effective and safe. Copyright © 2009 Informa UK Ltd.

Cite

CITATION STYLE

APA

Kuntsevich, V. I., Feinfeld, D. A., Audia, P. F., Young, W., Capponi, V., Markella, M., & Winchester, J. F. (2009). In-vitro myoglobin clearance by a novel sorbent system. Artificial Cells, Blood Substitutes, and Biotechnology, 37(1), 45–47. https://doi.org/10.1080/10731190802664379