Abstract
Loop-Mediated Isothermal Loop-Mediated Isothermal Amplification (LAMP) is a widely used technique for nucleic acid amplification due to its high specificity, sensitivity, and rapid results. Advances in microfluidic lab-on-chip (LOC) technology have enabled the integration of LAMP into miniaturized devices, known as (Formula presented.) -LAMP, which require precise thermal control for optimal DNA amplification. This paper introduces a novel thermal bed design using PCB copper traces and (Formula presented.) dielectric materials, providing a reliable, modular, and repairable heating platform. The system achieves accurate and stable temperature control, which is critical for (Formula presented.) -LAMP applications, with temperature deviations within ±1.0 °C. The thermal bed’s performance is validated through finite element method (FEM) simulations, showing uniform temperature distribution and a rapid thermal response of 2.5 s to reach the target temperature. These results highlight the system’s potential for applications such as disease diagnostics, biological safety, and forensic analysis, where precision and reliability are paramount.
Author supplied keywords
Cite
CITATION STYLE
Garcia-Torales, G., Torres-Ortega, H. H., Estrada-Marmolejo, R., Beltran-Gonzalez, A. B., & Strojnik, M. (2024). Thermal Bed Design for Temperature-Controlled DNA Amplification Using Optoelectronic Sensors. Sensors, 24(21). https://doi.org/10.3390/s24217050