MR-safe personal radiation dosimeters

9Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Magnetic resonance imaging (MRI) is being rapidly integrated for cancer treatments-such systems are referred to as MRI-guided radiation therapy (MRIgRT). As the magnet of an MRI scanner is always on, the presence of a strong static magnetic field from the MRI scanner during radiotherapy delivery presents new challenges. One of the challenges is that a personal radiation dosimeter used to estimate the radiation dose deposited in an individual wearing the device must be MR-safe. No such devices, however, are currently available. In this work we first modified an existing personal dosimeter (by removing a metal clip) to make it MRsafe and then investigated potential effects of magnetic field on dosimeter readings, i.e., optically stimulated luminescent dosimeter (OSLD) readings. We found that the effect of magnetic field on OSLD sensitivity was within radiation protection tolerance levels. OSLD personal dosimeters can be directly used in conjunction with MRIgRT radiation protection purposes.