The use of ELISA is comparable to immunoprecipitation in the detection of selected myositis-specific autoantibodies in a European population

11Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Background: The reliable detection of myositis-specific autoantibodies (MSA) provides valuable clinical information regarding prognosis, clinical progression and diagnostic confirmation. Objectives: To evaluate the reliability of a commercial ELISA immunoassay in detecting myositis-specific autoantibodies in comparison to immunoprecipitation as the reference standard. Methods: Serum samples were chosen from a biobank of more than 3000 samples. Samples with a confirmed MSA on Immunoprecipitation (n=116) were evaluated in duplicate by ELISA to detect Mi2, MDA5, Jo1, EJ, KS, PL-7 and PL-12 (Medical & Biological Laboratories Co. Ltd, Nagoya, Aichi, Japan). Healthy control samples (n=246) confirmed autoantibody negative by immunoprecipitation were similarly assessed. Results: There was a very good agreement between ELISA and immunoprecipitation for serum samples containing anti-Mi2, MDA5, Jo1, EJ, KS and PL-7 and PL-12 auto-antibodies. Cohen’s κ values ranged from 0.86-1 for the measured autoantibodies on ELISA. Conclusion: ELISA was an accurate method for detecting anti-synthetase, anti-Mi2 and anti-MDA5 autoantibodies.

Cite

CITATION STYLE

APA

Loganathan, A., McMorrow, F., Lu, H., Li, D., Mulhearn, B., McHugh, N. J., & Tansley, S. L. (2022). The use of ELISA is comparable to immunoprecipitation in the detection of selected myositis-specific autoantibodies in a European population. Frontiers in Immunology, 13. https://doi.org/10.3389/fimmu.2022.975939

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free