Safety and effectiveness of a novel dielectric mapping system: one-year, two chinese centers experiences

1Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: The KODEX-EPD system is a novel, dielectric three-dimensional mapping system. We aim to illustrate the feasibility, safety, and outcomes of ablation using the KODEX-EPD system. Methods: A total of 272 patients with supraventricular arrhythmias were enrolled and underwent catheter ablation using the KODEX-EPD system from October 2020 to July 2021. The feasibility, safety, and ablation outcomes were analyzed. Results: Of the enrolled patients, 15 (5.4%) had atrial tachycardia (AT), 88 (31.4%) had atrioventricular reentrant tachycardia (AVRT), 141 (50.4%) had atrioventricular nodal reentrant tachycardia (AVNRT), 34 (12.1%) had atrial fibrillation (AF), and 9 (3.2%) had atrial flutter (AFL). All AF patients included were first-do-pulmonary vein isolation (PVI); there were 26 paroxysmal AF and 8 persistent AF. All patients achieved immediate success of ablation. The mean follow-up duration was 11.8 ± 2.4 months. One patient (1.1%) in the AVRT subgroup and two patients (1.4%) in the AVNRT subgroup experienced recurrence. When considering a three-month blanking time, the estimated freedom of AF at one-year post-ablation with and without AADs was 75.7% and 70.4%, respectively. The Kaplan–Meier analysis showed no significant difference in the overall AF recurrence (log-rank; P = 0.931) or AAD-free AF recurrence (log-rank; P = 0.841) between RFCA and cryoablation. One patient had mild pulmonary embolism. None of the patients died or had a cerebrovascular event in the periprocedural period. Conclusions: This retrospective, two-center study demonstrated that catheter ablation of supraventricular arrhythmias using the KODEX-EPD system is feasible, safe, and effective. Trial registration Retrospectively registered.

Cite

CITATION STYLE

APA

Ding, L., Huang, X., Dai, C., Zhang, H., Weng, S., Yu, F., … Tang, M. (2022). Safety and effectiveness of a novel dielectric mapping system: one-year, two chinese centers experiences. BMC Cardiovascular Disorders, 22(1). https://doi.org/10.1186/s12872-022-02790-8

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free