Analysis of Focusing Performance in Phased Array Probe Using COMSOL

6Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

To improve the detection ability of ultrasonic phased array detection for hydrogen storage vessel, COMSOL was used to simulate the main parameters of phased array probes. The influence of the number of array elements, array element spacing and excitation frequency on the sound propagation characteristics was analysed. The results showed that the number of array elements has a significant effect on the focused sound field. When the number of array elements is 8, effective focus cannot be formed at the pre-set focus position, and the maximum sound pressure value does not change significantly at the three pre-set focus depths. At a focus depth of 30mm, the maximum sound pressure in the 32-element mode is more than twice that in the 8-element mode. The relationship between excitation frequency and focal sound pressure is non-linear. In addition, as the array element spacing increases, the length of the focal column decreases. And the smaller the array element spacing, the higher the sound pressure at a deeper depth.

Cite

CITATION STYLE

APA

Yan, R., & Zhou, C. (2023). Analysis of Focusing Performance in Phased Array Probe Using COMSOL. In Journal of Physics: Conference Series (Vol. 2610). Institute of Physics. https://doi.org/10.1088/1742-6596/2610/1/012059