Analysis of an automated background correction method for cardiovascular MR phase contrast imaging in children and young adults.

  • Rigsby C
  • Hilpipre N
  • McNeal G
 et al. 
  • 3

    Readers

    Mendeley users who have this article in their library.
  • N/A

    Citations

    Citations of this article.

Abstract

BACKGROUND: Phase contrast magnetic resonance imaging (MRI) is a powerful tool for evaluating vessel blood flow. Inherent errors in acquisition, such as phase offset, eddy currents and gradient field effects, can cause significant inaccuracies in flow parameters. These errors can be rectified with the use of background correction software.

OBJECTIVE: To evaluate the performance of an automated phase contrast MRI background phase correction method in children and young adults undergoing cardiac MR imaging.

MATERIALS AND METHODS: We conducted a retrospective review of patients undergoing routine clinical cardiac MRI including phase contrast MRI for flow quantification in the aorta (Ao) and main pulmonary artery (MPA). When phase contrast MRI of the right and left pulmonary arteries was also performed, these data were included. We excluded patients with known shunts and metallic implants causing visible MRI artifact and those with more than mild to moderate aortic or pulmonary stenosis. Phase contrast MRI of the Ao, mid MPA, proximal right pulmonary artery (RPA) and left pulmonary artery (LPA) using 2-D gradient echo Fast Low Angle SHot (FLASH) imaging was acquired during normal respiration with retrospective cardiac gating. Standard phase image reconstruction and the automatic spatially dependent background-phase-corrected reconstruction were performed on each phase contrast MRI dataset. Non-background-corrected and background-phase-corrected net flow, forward flow, regurgitant volume, regurgitant fraction, and vessel cardiac output were recorded for each vessel. We compared standard non-background-corrected and background-phase-corrected mean flow values for the Ao and MPA. The ratio of pulmonary to systemic blood flow (Qp:Qs) was calculated for the standard non-background and background-phase-corrected data and these values were compared to each other and for proximity to 1. In a subset of patients who also underwent phase contrast MRI of the MPA, RPA, and LPA a comparison was made between standard non-background-corrected and background-phase-corrected mean combined flow in the branch pulmonary arteries and MPA flow. All comparisons were performed using the Wilcoxon sign rank test (α = 0.05).

RESULTS: Eighty-five children and young adults (mean age 14 years; range 10 days to 32 years) met the criteria for inclusion. Background-phase-corrected mean flow values for the Ao and MPA were significantly lower than those for non-background-corrected standard Ao (P = 0.0004) and MPA flow values (P 
CONCLUSION: Our data suggest that background phase correction of phase contrast MRI data does not significantly change Qp:Qs quantification, and there are residual errors in expected Qp:Qs quantification despite background phase correction. However the use of background phase correction does improve quantification of MPA flow relative to combined RPA and LPA flow. Further work is needed to validate these findings in other patient populations, using other MRI units, and across vendors.

Get free article suggestions today

Mendeley saves you time finding and organizing research

Sign up here
Already have an account ?Sign in

Find this document

Authors

  • Cynthia K Rigsby

  • Nicholas Hilpipre

  • Gary R McNeal

  • Gang Zhang

  • Emma E Boylan

  • Andrada R Popescu

Cite this document

Choose a citation style from the tabs below

Save time finding and organizing research with Mendeley

Sign up for free