Comparative study of a novel biochemical assay, the rapidec carba NP test, for detecting carbapenemase-producing enterobacteriaceae

  • Lifshitz Z
  • Adler A
  • Carmeli Y
  • 12

    Readers

    Mendeley users who have this article in their library.
  • 16

    Citations

    Citations of this article.

Abstract

The novel biochemical test, the Rapidec Carba NP (RCNP), was evaluated using carbapenemase- and non-carbapenemase-producing Enterobacteriaceae isolates. The RCNP test was compared with the Carba NP test (CNP) and the modified Hodge test. Compared to the CNP test, the RCNP test had identical sensitivity (96%) and lower specificity (93% versus 100%). The medium used to culture the isolates significantly affected test sensitivity and specificity. The RCNP test was quicker and easier to perform than the other tests.

Get free article suggestions today

Mendeley saves you time finding and organizing research

Sign up here
Already have an account ?Sign in

Find this document

Authors

  • Ziv Lifshitz

  • Amos Adler

  • Yehuda Carmeli

Cite this document

Choose a citation style from the tabs below

Save time finding and organizing research with Mendeley

Sign up for free