Comparison of the Lactate Pro and Analox GM7 blood lactate analysers

  • Van Someren K
  • Howatson G
  • Nunan D
 et al. 
  • 24

    Readers

    Mendeley users who have this article in their library.
  • 6

    Citations

    Citations of this article.

Abstract

The purpose of this study was firstly, to determine the level of agreement between the Lactate Pro and Analox GM7 for the measurement of blood lactate, and secondly, to examine whether these analysers may be used interchangeably to identify lactate parameters routinely used in the physiological assessment of athletes. Twenty well-trained male cyclists performed an incremental cycle ergometry test; duplicate blood samples were taken simultaneously throughout the test for lactate determination using the two analysers. Power output and heart rate at LT, LT1, 2 mmol.L-1, and 4 mmol.L-1 were calculated from the lactate values obtained from the two analysers. There was a strong linear relationship between the two analysers (Lactate Pro=1.4541xAnalox GM7-0.1287; R2=0.969, p

Author-supplied keywords

  • Agreement
  • Blood lactate response
  • Lactate parameters
  • Training prescription