Comparison of RF body coils for MRI at 3T

  • Wu X
  • Zhang X
  • Tian J
  • et al.
ISSN: 0952-3480
N/ACitations
Citations of this article
2Readers
Mendeley users who have this article in their library.

Abstract

The performance of multichannel transmit coil layouts and parallel transmission (pTx) RF pulse design was evaluated with respect to transmit B 1 (B 1 +) homogeneity and specific absorption rate (SAR) at 3T for a whole body coil. Five specific coils were modeled and compared: a 32-rung birdcage body coil (driven either in a fixed quadrature mode or a two-channel transmit mode), two single-ring stripline arrays (with either 8 or 16 elements), and two multi-ring stripline arrays (with two or three identical rings, stacked in the z axis and each comprising eight azimuthally distributed elements). Three anatomical targets were considered, each defined by a 3D volume representative of a meaningful region of interest (ROI) in routine clinical applications. For a given anatomical target, global or local SAR controlled pTx pulses were designed to homogenize RF excitation within the ROI. At the B 1 + homogeneity achieved by the quadrature driven birdcage design, pTx pulses with multichannel transmit coils achieved up to about eightfold reduction in local and global SAR. When used for imaging head and cervical spine or imaging thoracic spine, the double-ring array outperformed all coils, including the single-ring arrays. While the advantage of the double-ring array became much less pronounced for pelvic imaging, with a substantially larger ROI, the pTx approach still provided significant gains over the quadrature birdcage coil. For all design scenarios, using the three-ring array did not necessarily improve the RF performance. Our results suggest that pTx pulses with multichannel transmit coils can reduce local and global SAR substantially for body coils while attaining improved B 1 + homogeneity, particularly for a "z-stacked" double-ring design with coil elements arranged on two transaxial rings.

Cite

CITATION STYLE

APA

Wu, X., Zhang, X., Tian, J., Schmitter, S., Hanna, B., Strupp, J., … de Moortele, P. F. V. (2015). Comparison of RF body coils for MRI at 3T. NMR in Biomedicine, 28(10), 1332–1344. Retrieved from http://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=84941740430&origin=inward

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free