Mechanics of stretchable electronics on balloon catheter under extreme deformation

  • Su Y
  • Liu Z
  • Wang S
 et al. 
  • 34

    Readers

    Mendeley users who have this article in their library.
  • 6

    Citations

    Citations of this article.

Abstract

Stretchable electronics has been applied to balloon catheters for high-efficacy ablation, with tactile sensing integrated on the surface, to establish full and conformal contact with the endocardial surface for elimination of the heart sink caused by blood flow around their surfaces. The balloon of the catheter folds into uniform 'clover' patterns driven by the pressure mismatch inside (∼vacuum) and outside of the balloon (pressure ∼1 atm). The balloon catheter, on which microelectrodes and interconnects are printed, undergoes extreme mechanical deformation during its inflation and deflation. An analytic solution is obtained for balloon catheter inflation and deflation, which gives analytically the distribution of curvatures and the maximum strain in the microelectrodes and interconnects. The analytic solution is validated by the finite element analysis. It also accounts for the effect of inflated radius, and is very useful to the optimal design of balloon catheter. © 2014 Published by Elsevier Ltd. All rights reserved.

Author-supplied keywords

  • Analytic model
  • Balloon catheter
  • Finite deformation
  • Stretchable electronics

Get free article suggestions today

Mendeley saves you time finding and organizing research

Sign up here
Already have an account ?Sign in

Find this document

Authors

  • Yewang Su

  • Zhuangjian Liu

  • Shuodao Wang

  • Roozbeh Ghaffari

  • Dae Hyeong Kim

  • Keh Chih Hwang

Cite this document

Choose a citation style from the tabs below

Save time finding and organizing research with Mendeley

Sign up for free