Abstract

OBJECTIVES: To prospectively evaluate dose reduction and image quality characteristics of chest CT reconstructed with model-based iterative reconstruction (MBIR) compared with adaptive statistical iterative reconstruction (ASIR).

METHODS: One hundred patients underwent reference-dose and low-dose unenhanced chest CT with 64-row multidetector CT. Images were reconstructed with 50 % ASIR-filtered back projection blending (ASIR50) for reference-dose CT, and with ASIR50 and MBIR for low-dose CT. Two radiologists assessed the images in a blinded manner for subjective image noise, artefacts and diagnostic acceptability. Objective image noise was measured in the lung parenchyma. Data were analysed using the sign test and pair-wise Student's t-test.

RESULTS: Compared with reference-dose CT, there was a 79.0 % decrease in dose-length product with low-dose CT. Low-dose MBIR images had significantly lower objective image noise (16.93 ± 3.00) than low-dose ASIR (49.24 ± 9.11, P < 0.01) and reference-dose ASIR images (24.93 ± 4.65, P < 0.01). Low-dose MBIR images were all diagnostically acceptable. Unique features of low-dose MBIR images included motion artefacts and pixellated blotchy appearances, which did not adversely affect diagnostic acceptability.

CONCLUSION: Diagnostically acceptable chest CT images acquired with nearly 80 % less radiation can be obtained using MBIR. MBIR shows greater potential than ASIR for providing diagnostically acceptable low-dose CT images without severely compromising image quality.

KEY POINTS: • Model-based iterative reconstruction (MBIR) creates high-quality low-dose CT images. • MBIR significantly improves image noise and artefacts over adaptive statistical iterative techniques. • MBIR shows greater potential than ASIR for diagnostically acceptable low-dose CT. • The prolonged processing time of MBIR may currently limit its routine use in clinical practice.

Author-supplied keywords

  • Adaptive statistical iterative reconstruction
  • Image noise
  • Model-based iterative reconstruction
  • Radiation dose reduction
  • Spatial resolution

Get free article suggestions today

Mendeley saves you time finding and organizing research

Sign up here
Already have an account ?Sign in

Find this document

Authors

  • Hiroyuki AkaiUniversity of Tokyo Institute of Medical Science

    Follow
  • Masaki Katsura

  • Izuru Matsuda

  • Masaaki Akahane

  • Jiro Sato

  • Koichiro Yasaka

Cite this document

Choose a citation style from the tabs below

Save time finding and organizing research with Mendeley

Sign up for free