Theoretical Calculation of Fluorescent X-Ray Intensities in Fluorescent X-Ray Spectrochemical Analysis.

  • Shiraiwa T
  • Fujino N
  • 24

    Readers

    Mendeley users who have this article in their library.
  • N/A

    Citations

    Citations of this article.

Abstract

The theoretical formulas of the fluorescent x-ray intensity (the primary, the secondary and the tertiary fluorescent x-rays) are obtained and actual calculations are made for nickel-iron, iron-chromium, and nickel-iron-chromium alloys. In the calculation, the continuous intensity distribution of the irradiating x-rays for the wavelength is taken into account. The theoretical results agree with the experiments very well, the deviations being 1% at most, and it is concluded that the theoretical estimation of the fluorescent x-ray intensity emitted from samples of various composition is possible. The matrix effect, absorption effect, and enhancement effect are evaluated from the present theory. In the enhancement effect, secondary fluorescent x-rays are quite significant (for instance, the chromium secondary fluorescent x-rays from Cr 2% and Fe 98% alloy reach the 55% of the chromium primary fluorescent x-rays and they can not be neglected), but the tertiary fluorescent x-rays are only a few per cent. The usual approximation methods, where the irradiating x-rays are assumed as monochromatic rays and the secondary fluorescent rays are not taken into consideration are also discussed and their meaning is made clear.

Get free article suggestions today

Mendeley saves you time finding and organizing research

Sign up here
Already have an account ?Sign in

Find this document

Authors

  • Toshio Shiraiwa

  • Nobukatsu Fujino

Cite this document

Choose a citation style from the tabs below

Save time finding and organizing research with Mendeley

Sign up for free