Enumeration of nucleated red blood cells with the ADVIA 2120 hematology system: An international multicenter clinical trial

52Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The accurate, reproducible, and timely reporting of nucleated red blood cells (NRBC) is an important function of the clinical hematology laboratory. We used 960 samples from 5 worldwide sites to evaluate a new NRBC enumeration method for the ADVIA 2120 Hematology System. The method showed excellent correlation with microscopy (r = 0.93). Sensitivity and specificity for the presence of NRBC for all samples analyzed was 77.3% and 74.6%, respectively. Almost all false negative samples were at NRBC counts ≤10/100 white blood cells (WBC). All false negative samples and the vast majority of the false positive specimens would be identified in clinical practice by routine slide review of flagged results, using existing review criteria. The coefficient of variation of the method was <10% for NRBC counts >10/100 WBC. The NRBC method automatically corrects WBC counts and differential results for the presence of NRBC; the uncorrected counts are available to the user on the run screen. This method allows the enumeration of NRBC with the ADVIA 2120 Hematology System in CBC/DIFF sample mode without need for additional hardware, sample preparation, or reagents. © 2006 Carden Jennings Publishing Co., Ltd.

Cite

CITATION STYLE

APA

Kratz, A., Maloum, K., O’Malley, C., Zini, G., Rocco, V., Zelmanovic, D., & Kling, G. (2006). Enumeration of nucleated red blood cells with the ADVIA 2120 hematology system: An international multicenter clinical trial. Laboratory Hematology, 12(2), 63–70. https://doi.org/10.1532/LH96.06010

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free