Objective: The present study aimed to investigate the dose differences and radiobiological assessment between Anisotropic Analytical Algorithm (AAA) and Acuros External Beam (AXB) with its 2 calculation models, namely, dose-to-water (AXB-Dw) and dose-to-medium (AXB-Dm), on esophageal carcinoma radiotherapy treatment plans. Materials and methods: The AXB-Dw and AXB-Dm plans were generated by recalculating the initial 66 AAA plans using the AXB algorithm with the same monitor units and beam parameters as those in the original plan. The dosimetric and radiobiological assessment parameters were calculated for the planning target volume (PTV) and organs at risk (OARs). The gamma agreement for the PTV and the correlation between it and the volume of the air cavity and bone among the different algorithms were compared simultaneously. The dose discrepancy between the theoretical calculation and treatment planning system (TPS) when switching from AXB-Dm to AXB-Dw was analyzed according to the composition of the structures. Results: The PTV dose of AXB-Dm plans was significantly smaller than that of the AAA and AXB-Dw plans (P
CITATION STYLE
Wang, L., Zhang, J., Huang, M., Xu, B., & Li, X. (2022). Radiobiological Comparison of Acuros External Beam and Anisotropic Analytical Algorithm on Esophageal Carcinoma Radiotherapy Treatment Plans. Dose-Response, 20(3). https://doi.org/10.1177/15593258221105678