This work describes the development of a home-made capillary electrophoresis (CE) system based on the capacitively coupled contactless conductivity detection (C4D) for the separation of the metallic species Zn2+, Cr3, Pb2, Cd2, Co 2, Cu2, Ni2 e TI+. A background electrolyte composed of MES/Histidine 0,02 mol L-1 (pH 5.0) was optimized for the separation of the metallic species by using organic solvents and complexing agents as additives. The system allowed the determination of the metallic species using MES/Histidine 0,02 mol L-1 and methanol 5% (pH 5.0) as a background electrolyte, 15 kV separation voltage and hydrodynamic injection by gravity.

Cite

CITATION STYLE

APA

Carvalho, L. M. de, Spengler, C., Lima, A. P. S. de, Raabe, A., Martini, M., Nascimento, P. C. do, & Bohrer, D. (2009). Determinação simultânea de cátions empregando eletroforese capilar com detecção condutométrica sem contato em equipamento construído em laboratório. Química Nova, 32(8), 2203–2208. https://doi.org/10.1590/s0100-40422009000800036