A novel immunoassay system based on the changes in the reflection of light, termed an optical immunoassay (OIA), was utilized to directly detect group A streptococcal (GAS) carbohydrate antigen from clinical specimens. In two studies, a total of 1,275 throat swabs were tested for the presence of this antigen with the Strep A OIA rapid detection system and the results were compared with those of standard culture methods. In both studies, the Strep A OIA yielded more positive results than plating of the throat swab onto a selective agar, Trypticase soy agar containing sheep blood, or an enriched broth. In one study, the sensitivity and specificity of Strep A OIA compared with those of the broth-enriched culture were 97.4 and 95.6%, respectively. In a second study a sensitivity of 98.9% and a specificity of 98.6% were achieved. It was also shown that the carbohydrate antigen could be detected in the absence of viable GAS organisms. The Strep A OIA is an easily interpretable method and was shown to be more sensitive than routine culture methods for detecting GAS infections directly from throat swabs.
CITATION STYLE
Harbeck, R. J., Teague, J., Crossen, G. R., Maul, D. M., & Childers, P. L. (1993). Novel, rapid optical immunoassay technique for detection of group A streptococci from pharyngeal specimens: Comparison with standard culture methods. Journal of Clinical Microbiology, 31(4), 839–844. https://doi.org/10.1128/jcm.31.4.839-844.1993
Mendeley helps you to discover research relevant for your work.