Polyethylene Glycol Functionalized Silicon Nanowire Field-Effect Transistor Biosensor for Glucose Detection

5Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

Abstract

Accurate monitoring of blood glucose levels is crucial for the diagnosis of diabetes patients. In this paper, we proposed a simple “mixed-catalyzer layer” modified silicon nanowire field-effect transistor biosensor that enabled direct detection of glucose with low-charge in high ionic strength solutions. A stable screening system was established to overcome Debye screening effect by forming a porous biopolymer layer with polyethylene glycol (PEG) modified on the surface of SiNW. The experimental results show that when the optimal ratio (APTMS:silane-PEG = 2:1) modified the surface of silicon nanowires, glucose oxidase can detect glucose in the concentration range of 10 nM to 10 mM. The sensitivity of the biosensor is calculated to be 0.47 μAcm−2mM−1, its fast response time not exceeding 8 s, and the detection limit is up to 10 nM. This glucose sensor has the advantages of high sensitivity, strong specificity and fast real-time response. Therefore, it has a potential clinical application prospect in disease diagnosis.

Cite

CITATION STYLE

APA

Zhu, Y., Wei, Q., Jin, Q., Li, G., Zhang, Q., Xiao, H., … Luo, Y. (2023). Polyethylene Glycol Functionalized Silicon Nanowire Field-Effect Transistor Biosensor for Glucose Detection. Nanomaterials, 13(3). https://doi.org/10.3390/nano13030604