Biomechanical Comparison of INTERTAN Nail and Gamma3 Nail for Intertrochanteric Fractures

19Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objective: To compare the biomechanical stabilities of Gamma3 nail and INTERTAN nail (ITN) for stable (AO/OTA 31A1.1) and unstable (AO/OTA 31A2.2) femoral intertrochanteric fracture. Methods: Twenty-four synthetic femora were randomly divided into four groups. After internal fixation (Gamma3 nail or ITN) had been implanted, stable and unstable intertrochanteric fracture models were produced. A cyclic testing protocol with increasing loads was performed for both stable and unstable intertrochanteric fracture models, and then torsional test and axial compression failure test were conducted. Stiffness, failure load, torque, and fragment displacement were recorded. Results: For stable fracture model: fragment displacement in ITN group were smaller than Gamma3 nail group (Gamma3 nail: 1.66 ± 0.13 mm; ITN: 1.55 ± 0.1 mm); stiffness (Gamma3 nail: 1142.6 ± 161.1 N/mm, ITN: 1159.3 ± 203.5 N/mm, P = 0.872) and failure load (Gamma3 nail: 5715.42 ± 616.34 N, ITN: 5690.27 ± 625.59 N, P = 0. 951) of the two nails were similar after cyclic test; torque of the ITN group was larger than the Gamma3 nail group. For unstable fracture model: fragment displacement in ITN group was significantly smaller than in the Gamma3 nail group when the axial load was larger than 800 N (Gamma3 nail: 3.59 ± 0.19 mm; ITN: 2.93 ± 0.28 mm); ITN group showed a significantly higher failure load than Gamma3 nail group (Gamma3 nail: 2942.77 ± 573.4 N, ITN: 3672.3 ± 790.5 N, P = 0. 011); torque was significantly higher for the ITN group compared to the Gamma3 nail group for three different angles. Conclusions: Both ITN and Gamma3 nail can maintain sufficient biomechanical stability for stable intertrochanteric fractures, but ITN was a better choice for unstable intertrochanteric fractures.

References Powered by Scopus

Epidemiology of hip fractures

635Citations
N/AReaders
Get full text

Nail or plate fixation of intertrochanteric hip fractures: Changing pattern of practice - A review of the American Board of Orthopaedic Surgery database

398Citations
N/AReaders
Get full text

Trochanteric Gamma nail and compression hip screw for trochanteric fractures: A randomized, prospective, comparative study in 210 elderly patients with a new design of the Gamma nail

233Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Finite-Element Analysis of a Novel Cephalomedullary Nail for Restricted Sliding to Reduce Risk of Implant Failure in Unstable Intertrochanteric Fractures

10Citations
N/AReaders
Get full text

Comparison of Cephalomedullary Nails with Sliding Hip Screws in Surgical Treatment of Intertrochanteric Fractures: A Cumulative Meta Analysis of Randomized Controlled Trials

8Citations
N/AReaders
Get full text

Dual-screw versus single-screw cephalomedullary nails for intertrochanteric femoral fractures: a systematic review and meta-analysis

7Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Luo, W., Fu, X., Ma, J. xiong, Huang, J. min, Wu, J., & Ma, X. long. (2020). Biomechanical Comparison of INTERTAN Nail and Gamma3 Nail for Intertrochanteric Fractures. Orthopaedic Surgery, 12(6), 1990–1997. https://doi.org/10.1111/os.12853

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 2

67%

Researcher 1

33%

Readers' Discipline

Tooltip

Medicine and Dentistry 4

80%

Engineering 1

20%

Save time finding and organizing research with Mendeley