Titer on chip: New analytical tool for influenza vaccine potency determination

20Citations
Citations of this article
31Readers
Mendeley users who have this article in their library.

Abstract

Titer on Chip (Flu-ToC) is a new technique for quantification of influenza hemagglutinin (HA) concentration. In order to evaluate the potential of this new technique, a comparison of Flu-ToC to more conventional methods was conducted using recombinant HA produced in a baculovirus expression system as a test case. Samples from current vaccine strains were collected from four different steps in the manufacturing process. A total of 19 samples were analysed by Flu-ToC (blinded), single radial immunodiffusion (SRID), an enzyme-linked immunosorbent assay (ELISA), and the purity adjusted bicinchoninic acid assay (paBCA). The results indicated reasonable linear correlation between Flu-ToC and SRID, ELISA, and paBCA, with regression slopes of log-log plots being 0.91, 1.03, and 0.91, respectively. The average ratio for HA content measured by Flu-ToC relative to SRID, ELISA, and paBCA was 83%, 147%, and 81%, respectively; indicating nearly equivalent potency determination for Flu-ToC relative to SRID and paBCA. These results, combined with demonstrated multiplexed analysis of all components within a quadrivalent formulation and robust response to HA strains over a wide time period, support the conclusion that Flu-ToC can be used as a reliable and time-saving alternative potency assay for influenza vaccines.

Cite

CITATION STYLE

APA

Kuck, L. R., Sorensen, M., Matthews, E., Srivastava, I., Cox, M. M. J., & Rowlen, K. L. (2014). Titer on chip: New analytical tool for influenza vaccine potency determination. PLoS ONE, 9(10). https://doi.org/10.1371/journal.pone.0109616

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free